DIASYS BIOTECHNOLOGY CO., Ltd.

Mail: contact@diasysbiotech.com

,

DIASYS DH 5100 5 Part Auto Hematology Analyzer

  • 3 Reagents
  • 3 Part & 5 part all-in one
  • 60 Samples/hour
  • Low consumption of reagent and sample Perfect combination of peripheral blood sampling and auto sample loading, It not only satisfies the sampling demand of children and elderly, but also does not affect the work efficiency.
Categories: , Tag:

Product Specifications:

Compatibility:

  • Two modes of sampling: Whole blood, pre-diluted blood
  • Three modes of test: CBC, CBC+5Diff
  • 4 USB, 1 LAN support protocol HL7 and compatible with LIS system
  • Accompanying quality control material
  • Auto-cleaning of sample probe and tubes
  • Automatic fault processing function

Principles:

  • Impedance method for RBC and PLT counting.
  • Cyanide free reagent for haemoglobin test.
  • Sheath Flow cytometry (FCM) + Tri-angle laser scatter method for WBC 5-part differential analysis and WBC counting.

Detection Parameter:

The detection items include Reportable Parameters (25 items) : WBC, NEU%, LYM%, MON%, EOS%, BASO%, NEU#, LYM#, MON#, EOS#, BASO#,, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC.

Research Parameters (12 items) : LIC%, LIC#, ALY%, ALY#, Blast%, Blast#, LIP%, LIP#, PLT Clumps%, PLT Clumps#, NRBC%, NRBC#

3 histograms for WBC, RBC and PLT.

3 DIFF scattergram and 1 BASO sacattergram.

Reagent:

  • Diluent, DIFF Lyse, LH Lyse for counting
  • Probe Cleaner for maintenance

Throughput:

60 Sample/Hour.

Sample Volume:

Minimum volume 20µL.

Repeatability:

WBC≤2.0%, RBC≤1.5%, HGB≤1.5%, PLT≤4.0%, MCV≤1.0%.

Linearity Range:

WBC (0.00-500.00)×10⁹/L

RBC (0.00-8.50)×10ˉ12/L

HGB (0-250)g/L

PLT (0-5000)×10⁹/L

HCT 0.0-67.0%

Storage:

Up to 50,000 results including numeric and graphical.

Printout:

Built-in thermal printer supports various types of printer. Multi-built in print templates available.

Display:

10.4 inch TFT touch screen.

Input/output:

4 USB ports and Bi-directional LIS connection.

Operating environment:

Temperature 10°C to 30°C; relative humidity ≤ 85%.

Power Requirement:

100V-240V~ ; 50/60Hz.

Dimensions:

420mm(L)×295mm(W)×430mm(H)

Net Weight:

24 KG.

Reviews

There are no reviews yet.

Be the first to review “DIASYS DH 5100 5 Part Auto Hematology Analyzer”